Search By Application
DEED (Dekati Engine Exhaust Diluter)
A simple, easy to use engine exhaust condtioning system that guarantees precise repeatable results every time.DEED Dekati Engine Exhaust Diluter
The DEED Dekati Engine Exhaust Diluter is an engine exhaust conditioning system which conforms to the new PMP legislation. In contrast to sample conditioning systems in general, DEED is a simple, easy to use system that guarantees precise and repeatable results every time.
The design criteria for the DEED were utmost simplicity, ease of use and reliability. The user interface only has two operation switches: high/low dilution ratio and heating on/off. The pre-set dilution ratios remain always constant, thus the measurement reproducibility is high. Furthermore, retention of particles inside DEED is minimal, thus leading to long maintenance intervals and minimum instrument downtime.
DEED fulfils all requirements and recommendations of the upcoming exhaust fine particle measurement legislation (PMP). The instrument can be controlled with AK-protocol and the entire system is contained in a single rack-mountable cabinet.
SYSTEM AND SPECIFICATIONS
Package includes
- PMP-compliant two-stage dilution system
- Integrated evaporation chamber
- Temperature controllers
- Hig/low dilution ratio selection
FEATURES
- Fulfils all PMP recommendations for sample conditioning
- Extremely low solid particle losses
- Long cleaning intervals
- Simple user interface – only two operation switches
- High/Low dilution ratio
- Heating on/off
- Always constant dilution ratio
- High reproducibility
- Dilution ratio calibrated with solid particles
- Remote AK-protocol control available
| Dilution ratio: | ~100 (with low DR selected) ~1000 (with high DR selected) |
|
| Fixed points can be adjusted upon request when ordering | ||
| Dilution ratio accuracy: | +/- 5% typical, +/-10% maximum | |
| Evaporation tube temperature: | 300-400 °C | |
| Evaporation efficiency: | >99% for 30nm tetracontane particles with inlet concentration above 10000 #/cm3 | |
| Particle concentration reduction factor: | As described in paragraph 1.333 in ECE/TRANS/WP.29/GRPE/2007/8/Rev.1 page 63 Appendix 5 | |
| Communication: | AK-protocol available | |
| Power supply: | 90-240 Volts, ~2 kW | |
| Pressurised air requirement: | 120 SLPM @ 3 bar gauge | |
| Dimensions: | 9U (40 cm) height, fits into 19" rack | |
| Sample intake: | Approx. 5 LPM | |
| Diluted sample output: | 0-40 LPM | |
Contact Biral for more information
To find out more about this instrument or to discuss your application requirements please do not hesitate to contact Biral.- wwwparticle@biral.com (or click the Make Request button below the product photograph)
- telephone to 01275 847787 (from the UK) or +44 1275 847787 (International).
After Sales Support
Biral offers free advice and technical support for the lifetime of the instruments we sell. Once you have purchased the unit you can contact us for help:- wwwaerosolsupport@biral.com
- telephone to 01275 847787 (from the UK) or +44 1275 847787 (International).
Product Information
Product Manuals
Application Related Information
FINE PARTICLE SAMPLER
FPS-4000
The most comprehensive dilution and sample conditioning system commercially available in the world.
From £27,693.00
From £27,693.00
THERMODENUDER FOR REMOVAL OF VOLATILES 110V
ELA-111
The ELA-111 Dekati Thermodenuder is an easy-to-use and simple tool for removing volatile and semi-volatile compounds from automotive exhaust.
£9,213.00
£9,213.00




