Search By Application
HYGRO-THERMO TRANSMITTER (0 - 1V)
This instrument is designed for the measurement of temperature and air humidity with the data output as electrical analogue signals. The transmitters consist of a capacitive humidity element and a Pt 100 resistance thermometer. For mounting outdoors it must be used with the weather and thermal radiation shield part number 1.1025.55.xxxCompact Hygro Thermo Transmitter
(model 1.1005.54.xxx)
The Compact Hygro Thermo Transmitter is designed to measure relative humidity and temperature of the air and other non-aggressive gases. It is supplied with a 5 m connecting cable.
The use of capacitive humidity sensors guarantees:
- a high degree of long-term stability
- nearly linear characteristics
- good dynamic behaviour
- dewing stability
- low temperature coefficients
- low hysteresis
| NOTE for use outside it is advisable to use a Weather and Thermal Radiation Shield.... which is available as an optional accessory. |
Maintenance
The Hygro Thermo transmitter is delivered already calibrated and its characteristics remain stable for years.Dust does not damage the humidity sensor but does negatively affect the dynamic behaviour. If the instrument is very dirty, the sensor element can be cleaned or carefully rinsed in distilled water making sure that the highly sensitive element is not touched. Before cleaning the sensor elements the protective filter should be cleaned or replaced.
There are a wide choice of models available and below are the most popular ones. Please see the manual in the downloads tab for other models and contact us for a price.
| Order-No | Electrical Output | Operating voltage |
| 1.1005.54.000 | Humidity: 0 - 1 V Temperature: Pt 100 | 6 - 30 VDC |
| 1.1005.54.161 | Humidity: 0 - 10 V Temperature: 0 - 10 V | 15 - 30 VDC |
| 1.1005.54.241 | Humidity: 4 - 20 mA Temperature: 4 - 20 mA | 12 - 30 VDC |
|
SPECIFICATIONS for models 1.1005.54.000, 1.1005.54.161 and 1.1005.54.241 |
|
| Humidity | |
| Measuring Element | Capacitive |
| Measuring Range | 0 - 100% relative humidity |
| Deviation (mr 5 - 95 % rel humidity at 10 - 40 °C) |
± 2% relative humidity |
| Add. error (<10 °C, >40 °C) | <0.1 %/K |
| Response Time (T 90) | < 20 s (at v = 1.5 m/s) without filter < 1.5 min (at v = 1.5 m/s) with Teflon filter < 1.5 min (at v = 1.5 m/s) with Sinter filter |
| Temperature | |
| Measuring Element | Pt 100 Class B, 1/3 DIN tolerance |
| Measuring range | <-30 to +70 °C |
| Deviation | |
| with outputs 0-1 V and -0 - 10 V | ± 0.2 K |
| with output 4 - 20 mA | ± 0.3 K |
| Add. error (<10 °C, >40 °C) | ± 0.0073 K/K |
| Response Time (T 90) | < 20 s (at v = 1.5 m/s) without filter < 1.5 min (at v = 1.5 m/s) with Teflon filter < 1.5 min (at v = 1.5 m/s) with Sinter filter |
| Additional specifications | |
| Ambient temperature | -40 to +80 °C -40 to +100 °C (model ...000) |
| Degree of protection - sensor | IP 30 |
| Degree of protection - electronics, connecting head | IP 65 |
| Operating voltage | |
| I-Output | 12 ... 30 V DC |
| U- Output (0..10 V) | 15 ... 30 V DC |
| U- Output (0.. 1 V) | 6 ... 30 V DC |
| Load Resistor | |
| I- Output | see diagram RL in manual (download tab) |
| U- Output (0 -10 V) | _/> 10 k Ohms |
| U- Output (0 -1 V) | _/> 2 kOhms |
| Current requirements | |
| Humidity / temperature (0 - 10 V) | < 5 mA |
| Humidity (0 - 1V) | <1 mA |
| Dimensions (for models ...000 and ...161) | Ø 20 mm |
| Shaft length | 122 mm |
| Total length | 145 mm |
You can order this instrument on-line
If you would like to place an order please click the Make Request button and check the box for Do you want to order this product now? in the Send Info form.We will then contact you to discuss your preferred payment method.
Please see our Terms and Conditions
Any questions?
If you have any questions or wish to discuss this sensor's suitability for your application please do not hesitate to contact Biral.- wwwmet@biral.com
- telephone to 01275 847787 (from the UK) or +44 1275 847787 (International)